AH-800 Online Water Hardness and Alkali Analyzer

Online Water Hardness/Alkali Analyzer
Model:AH-800



Online Water hardness and Alkali analyzer monits water total hardness or carbonate hardness and total alkali fully automatically via titration.

Description
This analyzer can measure water total hardness or carbonate hardness and total alkali fully automaticlly via titration. This instrument is suitable for recognising levels of hardness, quality control of water softening facilities and monitoring of water blending facilities. The instrument allows two different limit values to be defined and checks the water quality by determining the absorbtion of the sample during titration of the reagent. The configuration of the many applications is supported by a configuration assistant.

Features
◇ Reliable, exact and fully automatic analysing
◇ Simple commissioning with configuration assistant
◇ Self calibrating and self monitoring
◇ High measuring accuracy
◇ Easy maintenance and cleaning.
◇ Minimal reagent and water consumption
◇ Multi-coloured and multi-lingual graphic display.
◇ 0/4-20mA/relay/CAN-interface output



Hardness Reagents & Measurement Ranges
Reagent type
°dH
°F
ppm CaCO3
mmol/l
TH5001
0.03-0.3
0.053-0.534 
0.534-5.340 
0.005-0.053
TH5003
0.09-0.9
0.160-1.602
1.602-16.02
0.016-0.160
TH5010
0.3-3.0
0.534-5.340
5.340-53.40
0.053-0.535
TH5030
0.9-9.0
1.602-16.02
16.02-160.2
0.160-1.602
TH5050
1.5-15
2.67-26.7
26.7-267.0
0.267-2.670
TH5100
3.0-30
5.340-53.40
53.40-534.0
0.535-5.340


Alkali Reagents & Measurement Ranges
Reagents model
Measuring rang
TC5010
5.34~134 ppm
TC5015
8.01~205ppm
TC5020
10.7~267ppm
TC5030
16.0~401ppm


Specifications
Measurement method
Titration method
Water inlet in general
clear, colourless, free of solid particles, without gas bubbles
Measurement range
Hardness : 0.5-534ppm , total alkali:5.34~401ppm
Accuracy
+/- 5% 
Repetition
±2.5%
Environmental temp.
5-45℃
Measuring water temp.
5-45℃
Water inlet pressure
ca. 0.5 - 5 bar (max.) (Recommended 1 - 2 bar)
Analysis start
- programmable time intervals (5 - 360 minutes)
- external signal
- programmable volume intervals
Flush time
programmable flush time (15 - 1800 seconds)
Output
- 4 x potential-free Relays (max. 250 Vac / Vdc; 4Aas potential free output NC/NO)
- 0/4-20mA
- CAN interface
Power
90 - 260 Vac (47 - 63Hz)
Power consumption
25 VA (in operation), 3.5 VA (stand by)
Dimensions
300x300x200 mm (WxHxD)
Protection grade
IP65