DDG-3080 Industrial TDS Salinity ppm Meter
DDG-3080 Industrial TDS Salinity/ppm
Meter
DDG-3080 Industrial TDS Salinity ppm Meter is
our company’s latest top-class instrument equipped with a microprocessor and
complete English menu. It has display and menu operation pattern; it is
characterized by its intelligent operation, multi-function and high measuring
performance and its strong environmental adaptability. It can be widely used
for continuous monitoring of the TDS of solutions used in the following fields:
thermal power plant, chemical fertilizer, chemical industry, metallurgy,
environmental protection, pharmacy manufacturing, biochemistry, foodstuff and
running water.
Features of DDG-3080 Industrial TDS Salinity ppm Meter
|
u It has a complete English
display and friendly interface. Various parameters can be displayed at the
same time: TDS, Salinity, output current, temperature, time and status.
Bitmap type liquid crystal crystal display module.
|
|
|
|
u With high resolution
is adopted. All the data, status and operation prompts are displayed in
English. There is no symbol or code that is defined by the manufacturer.
|
|
|
Technical Indexes of DDG-3080 Industrial TDS Salinity ppm Meter
|
1. Measuring range: 0-100ppt
|
|
2. Intrinsic error
of the electronic unit: TDS: ±0.5%FS, temperature: ±0.3℃
|
|
3. Range of automatic
temperature compensation: 0~199.9℃, with 25℃ as the reference temperature.
|
|
4. Water sample
tested: 0~199.9℃, 0.6MPa
|
|
5. Intrinsic error of the instrument: TDS: ±1.0%FS, temperature:
±0.5℃
|
|
6. Automatic
temperature compensation error of the electronic unit: ±0.5%FS
|
|
7. Repeatability
error of the electronic unit: ±0.2%FS±1 Unit
|
|
8. Stability of the
electronic unit: ±0.2%FS±1 unit/24h
|
|
9. Isolated current
output: 0~10mA ( load<1.5kΩ); 4~20mA (load<750Ω) (double-current output for option)
|
10. Output current error: ≤±l%FS |
|
11. Error of
electronic unit caused by ambient temperature: ≤±0.5%FS
|
|
12. Error of the
electronic unit caused by the supply voltage: ≤±0.3%FS
|
|
13. Alarm relay: AC
220V, 3A
|
|
14. Communication
interface: RS485 or 232 (optional)
|
|
15. Power supply: AC
220V±22V, 50Hz±1Hz, 24VDC (optional,or personal design)
|
|
16. Protection
grade: IP54 , aluminium shell suitable for outdoor use
|
|
17. Clock accuracy:
±1 minute/month
|
|
18. Data storage
capacity: l month (1 point/5 minutes)
|
|
19. Saving time of
data under a continuous power-failure condition: 10 years
|
|
20. Overall dimension:
146 (length) x 146 (width) x 150 (depth) mm; dimension of the hole: 138 x
138mm
|
|
21. Working
conditions: ambient temperature: 0~60℃; relative humidity <85%
|
|
22. Weight: 1.5kg
|
|
23. The conductivity
electrodes with the following five constants are usable: K=0.01, 0.1, 1.0,
10.0, and 30.0.
|
DDG-3080 Industrial TDS Salinity ppm Meter series of electrodes are specially used for the measurement of conductivity value of pure water, ultra-pure water, water treatment, etc. It is especially suitable for conductivity measurement in the thermal power plant and the water treatment industry. It is featured by the double-cylinder structure and the titanium alloy material, which can be naturally oxidized to form the chemical passivation. Its anti-infiltration conductive surface is resistant to all kinds of liquid except fluoride acid. The temperature compensation components are: NTC2.252K, 2K, 10K, 20K, 30K, ptl00, ptl000, etc. which are specified by the user. K=10.0 or K=30 electrode adopts a large area of platinum structure, which is resistant to strong acid and alkaline and has strong anti-pollution capacity; it is mainly used for on-line measurement of the conductivity value in the special industries, such as the sewage treatment industry and the seawater purification industry.
Technical
Indexes of DDG-3080 Industrial TDS Salinity ppm Meter
1. Constant of electrode: 10.0
2. Compressive strength: 0.6MPa
3. Measuring range: 0-100ppt
4. Connection: 1/2or 3/4 Thread Installation
5. Material: polysulfone
6. Application: Water Treatment Industry
if any qeustions about DDG-3080 Industrial TDS Salinity ppm Meter , ,pls contact email : michael@shboqu.com