DOG-3082 Industrial Dissolved Oxygen Meter
DOG-3082
Industrial Dissolved Oxygen Meter
DOG-3082 Industrial
online Dissolved Oxygen Meter is our latest generation
of microprocessor-based
high-intelligence on line
meter, with English display, menu operation, a high
intelligent, multi-function, high measurement performance, environmental
adaptability and other characteristics, used for continuous on line
monitoring. It can be equipped with DOG-208F Polarographic Electrode and can
automatically switch from ppb level to ppm level of wide-range measurement.
This instrument is designed for monitoring the oxygen content in the boiler
feed water, condensate water and sewage.
Features of DOG-3082 Industrial Dissolved Oxygen Meter
New design, Aluminium shell, Metal
texture.
|
All the data are
displayed in English. It can be easily operated:
|
|
It has a complete
English display and elegant interface: Liquid crystal display module with
high resolution is
|
|
adopted. All the data, status and operation prompts are
displayed in English. There is no symbol or code that is
|
|
defined by the
manufacturer.
|
|
Simple menu structure and text-type man-instrument
interaction: Compared with traditional instruments,
|
|
DOG-3082 has many
new functions. As it adopts classified menu structure, which is similar to
that of a computer,
|
|
it is clearer and more convenient. It is not necessary to
remember the operation procedures and sequences. It can
|
|
be operated
according to the prompts on the screen without the guidance of an operation
manual.
|
|
Multi-parameter display: The oxygen concentration value,
input current (or output current), temperature values,
|
|
time and status can
be displayed on the screen at the same time. The main display can show the
oxygen
|
|
concentration value in a 10 x 10mm size. As the main
display is eye-catching, the displayed values can be seen
|
|
from a long distance
away. Six sub-displays can display such information as the input or output
current,
|
|
temperature, status, week, year, day, hour, minute and
second, so as to adapt to different users’ habits and to
|
|
conform with
different reference times set by the users.
|
Technical Indexes of DOG-3082 Industrial Dissolved Oxygen Meter
|
1. Measuring range:
0~100.0ug/L; 0~20.00 mg/L
(automatic switching); (0-60℃);(0-150℃)Option
|
|
Resolution: 0.1ug/L;
0.01 mg/L; 0.1℃
|
|
2. Intrinsic error
of the whole instrument: ug/L: ±l.0%FS; mg/L: ±0.5%FS, temperature: ±0.5℃
|
|
3. Repeatability of
the indication of the whole instrument: ±0.5%FS
|
|
4. Stability of the
indication of the whole instrument: ±1.0%FS
|
|
5. Automatic
temperature compensation range: 0~60℃, with 25℃ as the reference temperature.
|
|
6. Response time:
<60s (98% and 25℃ of the final value) 37℃: 98% of the final value < 20
s
|
|
7. Clock accuracy:
±1 minute/month
|
|
8. Output current
error: ≤±l.0%FS
|
|
9. Isolated output:
0-10mA (load resistance <1.5KΩ); 4-20mA (load resistance <750Ω)
|
|
10. Communication
interface: RS485 (optional) (Double power for option)
|
|
11. Data storage
capacity: l month (1 point/5 minutes)
|
|
12. Saving time of
data under a continuous power-failure condition: 10 years
|
|
13. Alarm relay: AC
220V, 3A
|
|
14. Power supply:
220V±10% 50±1HZ, 24VDC(option)
|
|
15.Protection: IP54,
Aluminium shell
|
|
16. Size: secondary
meter: 146 (length) x 146 (width) x 150(depth) mm;
|
|
dimension of the
hole: 138 x 138mm
|
|
17. Weight: 1.5kg
|
|
18. Working
conditions: ambient temperature: 0-60℃; relative humidity <85%
|
|
19. The connection
tubes for inlet and outlet water: Pipes and hoses
pls contact email: michael@shboqu.com; if any questions of DOG-3082 Industrial Dissolved Oxygen Meter
|

评论