Digital RS485 Modbus or 4-20mA output Residual Chlorine Sensor
Digital RS485 Modbus or 4-20mA output Residual Chlorine Sensor
New Residual Chlorine Sensor,
it can be directly with RS485 Modbus or 4-20mA output.max pressure is 3bar
indicator
|
Free chlorine
reduced dependence on pH
|
Application
|
e. g. Swimming pool water, drinking water, sea water Surfactants (tensides) are partially tolerated.
|
Chlorination agents
|
inorganic chlorine compounds:
NaOCl (=sodium hypochlorite), Ca(OCl)2, chlorine gas, electrolytically generated chlorine
|
Measuring system
|
Membrane covered, amperometric potentiostatic 3-electrode system with electronic inside
|
Electronic
|
Analog version: - voltage output
- not galvanically isolated electronics
- analog internal data processing
- output signal: analog (analog-out/analog) Digital version: - electronic is completely galvanically isolated
- digital internal data processing
- output signal: analog (analog-out/digital)
or
digital (digital-out/digital) mA-version: - current output analog
- not galvanically isolated electronics
- output signal: analog (analog-out/analog)
|
Information about the measuring range of sensors with 4-20 mA
|
Slope of a sensor can vary production-related or application-related between 65% and 150% of the nominal slope
-> Recommendation to determine the suitable measuring range or the suitable sensor:
Concentration to be measured x factor 1.5 = measuring range of the sensor
Example: Concentration to be measured 1.6 ppm x 1.5 = 2.4
-> recommended sensor with a measuring range of 5 ppm
|
Accuracy
after calibration at repeatability conditions (25°C, pH 7.2 in drinking water) of the upper full scale
|
- Measuring range 2 mg/l: at 0.4 mg/l <1%
at 1.6 mg/l <1%
- Measuring range 20 mg/l: at 4 mg/l <1%
at 16 mg/l <3%
|
Slope drift
At repeatability conditions (25 °C, pH 7,2 in drinking water)
|
approx. -1% per month
|
Working temperature
|
Measuring water temperature: 0 … +45 °C
(no ice crystals in the measuring water)
|
Ambient temperature: 0 … +55 °C
| |
Temperature compensation
|
Automatically, by an integrated temperature sensor Sudden temperature changes must be avoided
|