PHG-2518 Industrial PH Transmitter
PHG-2518
Industrial PH Transmitter
PHG-2518 Industrial online PH
Transmitter is our company’s latest-developed two (four)-wire transmitter type
intelligent integrated industrial PH Transmitter, which adopts the latest
design pattern and is characterized by its high performance and reliability as
well as LCD display. This meter is specially developed for measuring the PH
values of water used in such industries as chemical industry, power stations,
electronic industry and fermentation industries.
PHG-2518 industrial PH transmitter
is a kind of industrial PH meter for continuous measuring. Its pre-amplifier is
a high-quality high-impedance amplifier. It used 3-in-1 PH electrodes. This
meter is characterized by high accuracy, fast response, small volume and easy
installation and maintenance.
This instrument adopts two-point
calibration and a two-wire system with 4~20mA output or a four-wire system with
24V power supply and 4~20mA output. As the instrument transmits the PH signals
by means of electric current, it possesses strong interference immunity and the
capacity of long-distance transmission that can reach as far as 200 meters.
Features of PHG-2518 Industrial PH Transmitter
Equipped with the latest single
chip microcomputer, this PH meter is characterized by its low power
|
consumption, less peripheral
components and high performance, high stability and reliability.
|
This instrument adopts LCD display. When you press the
keys, the PH meter can display the PH values, mV
|
value and temperature values
respectively.
|
The meter has
power-failure parameter protection.
|
It has two-point or one-point
automatic calibration function that is convenient and accurate.
|
Automatic
temperature compensation.
|
It can display the
“zero-potential” pH value and slope of the electrode system.
|
It adopts 4~20mA current
output.
|
Technical Indexes of PHG-2518 Industrial PH Transmitter
1. Working
conditions
|
a. Ambient temperature: 0~55℃
|
b. Relative humidity<85% >
|
c. Power supply: 24 VDC±5%
|
d. Solution temperature tested:
5~55℃
|
2. Measuring range
|
PH: 0~14PH
|
Temperature: 0~85℃
|
3. Intrinsic errors
|
PH: ±0.05PH
|
Temperature: ±0.5℃
|
4. Input impedance:>l012Ω
|
5. Temperature
compensation range and influence error: 5~55℃ ±0.5%FS/10℃
|
6. Current output: 4~20mA (RL≤500Ω) ±0.5%FS.
pls contact email: michael@shboqu.com; if any questions of PHG-2518 Industrial PH Transmitter
|